Abstract

The chromatographic performance of 50 μm internal diameter (i.d.) fused silica columns coated with up to 2 μm films of immobilized SE‐54 (methyl phenyl (5%) silicone) is evaluated under gas chromatographic conditions. The influence of pressure drop on the plate height is discussed. In comparison to thick film 250–530 μm i.d. columns, much higher efficiencies and faster analyses are obtained. Practical examples, performed on a standard GC instrument, illustrate the features of thick film 50 μm i.d. columns.