๐”– Bobbio Scriptorium
โœฆ   LIBER   โœฆ

Capacitance based mass metering for cryogenic fluids

โœ Scribed by Mark A. Nurge; Robert Youngquist; Deron Walters


Publisher
Elsevier Science
Year
2003
Tongue
English
Weight
320 KB
Volume
43
Category
Article
ISSN
0011-2275

No coin nor oath required. For personal study only.

โœฆ Synopsis


This paper describes a method for measuring the mass of cryogenic fluids in on-board rocket propellant tanks or ground storage tanks. Linear approximations to the Clausius-Mossotti relationship serve as the foundation for a capacitance based mass sensor, regardless of fluid density stratification or tank shape. Sensor design considerations are presented along with the experimental results for a capacitance based mass gage tested in liquid nitrogen. This test data is shown to be consistent with theory resulting in a demonstrated mass measurement accuracy of ยฑ0.75% and a deviation from linearity of less than ยฑ0.30% of full scale mass. Theoretical accuracies are also shown to be ยฑ0.73% for hydrogen and ยฑ1.39% for oxygen for a select range of pressures and temperatures.