Can you trust your cryostat? Reproducibility of cryostat section thickness

✍ Scribed by Philip C. De Witt Hamer; Fonnet E. Bleeker; Aeilko H. Zwinderman; Cornelis J.F. Van Noorden


Book ID
102333180
Publisher
John Wiley and Sons
Year
2006
Tongue
English
Weight
96 KB
Volume
69
Category
Article
ISSN
1059-910X

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Reproducibility of cryostat section thickness is required for valid quantitative microscopy. This is generally pursued by motorized sectioning using a low but constant speed. The purpose of our study was to compare variation in section thickness between motorized and manual cryostat sectioning. Serial sections were cut from a frozen block of homogenized tissue on different days. Lactate dehydrogenase activity was histochemically detected and calibrated absorbance measurements were taken. The coefficients of variation of measurements was 9.7% for motorized sectioning and 3.3% for manual sectioning. In conclusion, section thickness is similarly reproducible after manual sectioning compared with motorized sectioning, if not better. Microsc. Res. Tech., 2006. © 2006 Wiley‐Liss, Inc.