Calibration of the radio frequency field for magnetic resonance imaging
โ Scribed by David L. Foxall; Bernice E. Hoppel; Hari Hariharan
- Publisher
- John Wiley and Sons
- Year
- 1996
- Tongue
- English
- Weight
- 640 KB
- Volume
- 35
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
โฆ Synopsis
Abstract
We have developed and validated the performance of a novel slice selective pulse sequence that allows direct calibration of the RF field using a simple rectangular pulse. The new sequence offers a number of substantial advantages. It operates at steady state and has an accurate calibration response at short repetition times. The slice selection train is insensitive to RF field strength changes caused by patient loading. The issue of patient motion has been addressed in our data collection and analysis routines. The applicability of the method to human scanning has been demonstrated in the automated RF power calibration routine of a commercial imaging system.