𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Calibration of the pressure dependence of the R1 ruby fluorescence line to 195 kbar

✍ Scribed by Piermarini, G. J.; Block, S.; Barnett, J. D.; Forman, R. A.


Book ID
115533309
Publisher
American Institute of Physics
Year
1975
Tongue
English
Weight
840 KB
Volume
46
Category
Article
ISSN
0021-8979

No coin nor oath required. For personal study only.

✦ Synopsis


The pressure dependence of the R1 ruby fluorescence line has been calibrated at 25 °C against the compression of NaCl. Pressures are determined using the Decker equation of state for NaCl. The dependence is linear to 195 kbar following the equation PNaCl=2.746(Δλ), where P is in kbar and Δλ in Γ…. The uncertainty in the value of the slope, dp/dΞ», expressed in terms of a 95% confidence interval is 2.746Β±0.014 kbarβ€‰Γ…βˆ’1. The coefficient of the quadratic term (Δλ)2 is not significantly different from zero; and the quadratic term makes indeed a negligible contribution to the fit. Taking into account the reported uncertainty associated with the Decker equation of state for NaCl, the value of the slope is 2.740Β±0.016 kbarβ€‰Γ…βˆ’1 within a 95% confidence interval.