Calibration of the pressure dependence of the R1 ruby fluorescence line to 195 kbar
β Scribed by Piermarini, G. J.; Block, S.; Barnett, J. D.; Forman, R. A.
- Book ID
- 115533309
- Publisher
- American Institute of Physics
- Year
- 1975
- Tongue
- English
- Weight
- 840 KB
- Volume
- 46
- Category
- Article
- ISSN
- 0021-8979
- DOI
- 10.1063/1.321957
No coin nor oath required. For personal study only.
β¦ Synopsis
The pressure dependence of the R1 ruby fluorescence line has been calibrated at 25βΒ°C against the compression of NaCl. Pressures are determined using the Decker equation of state for NaCl. The dependence is linear to 195 kbar following the equation PNaCl=2.746(ΞΞ»), where P is in kbar and ΞΞ» in Γ . The uncertainty in the value of the slope, dp/dΞ», expressed in terms of a 95% confidence interval is 2.746Β±0.014 kbarβΓ β1. The coefficient of the quadratic term (ΞΞ»)2 is not significantly different from zero; and the quadratic term makes indeed a negligible contribution to the fit. Taking into account the reported uncertainty associated with the Decker equation of state for NaCl, the value of the slope is 2.740Β±0.016 kbarβΓ β1 within a 95% confidence interval.