Calibration of reaction temperatures in a very high pressure shock tube using chemical thermometers
✍ Scribed by R. S. Tranter; R. Sivaramakrishnan; N. Srinivasan; K. Brezinsky
- Publisher
- John Wiley and Sons
- Year
- 2001
- Tongue
- English
- Weight
- 154 KB
- Volume
- 33
- Category
- Article
- ISSN
- 0538-8066
- DOI
- 10.1002/kin.1069
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Two chemical thermometers, 1,1,1‐trifluoroethane and cyclohexene, have been used to calibrate the temperatures behind reflected shock waves, T~5real~, in a unique high‐pressure, single‐pulse shock tube. Experiments with 1,1,1‐trifluoroethane were performed at nominal postshock pressures of 5000 psi and 9000 psi, and T~5real~ was calculated from the extent of the reaction and literature values for k~∞~. Both parent molecule decomposition and product formation were used to calculate the extent of reaction. At each pressure the two methods of calculating T~5real~ are in excellent agreement. The values for T~5real~ obtained from the cyclohexene experiments are in good agreement with those from 1,1,1‐trifluoroethane. The range of the temprature calibration is 1050–1350 K. No discernable pressure dependence between the two sets of experiments was observed for the calculated values of T~5real~. The effect of a 10% rise in pressure during the residence time on the calculated temperatures was also examined and is found to give rise to only small errors in T~5real~. © 2001 John Wiley & Sons, Inc. Int J Chem Kinet 33: 722–731, 2001