## Abstract The portable surface capacitance insertion probe was used to measure the __in situ__ water content of near surface soil layers. The probe readings were calibrated against gravimetric samples collected over a wide range of water contents, and were found to be very closely correlated. The
Calibration of a TDR probe for low soil water content measurements
β Scribed by Wojciech Skierucha; Andrzej Wilczek; Olga Alokhina
- Publisher
- Elsevier Science
- Year
- 2008
- Tongue
- English
- Weight
- 750 KB
- Volume
- 147
- Category
- Article
- ISSN
- 0924-4247
No coin nor oath required. For personal study only.
β¦ Synopsis
Time domain reflectometry (TDR) probes are increasingly used for field and laboratory estimation of soil water content. Usual calibration of TDR probes for the determination of soil water content uses two media: air for low and water for high values of dielectric permittivity, although the measured range of dielectric permittivity in soil is much smaller as compared to the range implied by the calibration media. The use of air for calibration of short TDR probes gives calibration errors due to overlapping incident and reflected pulses in the reflectogram, which result in their relative shift in time. This phenomenon, named the convolution effect, can be avoided by the application of selected calibration media. The presented approach minimizes of dielectric permittivity measurement errors by choosing the calibration media with dielectric permittivity values close to the limits of the measurement range and the possibility to use TDR probes of various lengths. The comparison of errors of TDR apparent dielectric permittivity measurement in three sandy soils, based on the probe calibrations in various media, is also presented.
π SIMILAR VOLUMES