Buffer composition suitable for determining very low fluoride concentrations using a fluoride ion-selective electrode and its application to the continuous analysis of rain water

✍ Scribed by Hirokazu Hara; Chun-ching Huang


Publisher
Elsevier Science
Year
1997
Tongue
English
Weight
558 KB
Volume
338
Category
Article
ISSN
0003-2670

No coin nor oath required. For personal study only.

✦ Synopsis


In order to continuously analyze rain water with a fluoride ion-selective electrode (F-ISE), various buffer systems were examined in terms of the sensitivity enhancement. On the basis of the result that the sensitivity could be maximized at around pH 2.8 using the conventional acetate buffer, we proposed Strrensen's buffer (a mixture of glycine and hydrochloric acid, pH 2.8) to be most suitable for rain water analysis. The limit of linear response was 1 x lo-' mol dmP3 and the detection limit was I x lo-' mol dm-' by optimizing the sample to buffer dilution rate as 10 : 1. A good correlation was found between the results from the rain water analysis using this method and those of ion chromatography. A continuous analysis of rain water during a rainfall event was successfully carried out.