Balloon atrial septostomy using a new low-profile balloon catheter: Initial clinical results

✍ Scribed by Hijazi, Ziyad M. ;Ata, Ibrahim Abu ;Kuhn, Micheal A. ;Cheatham, John P. ;Latson, Larry ;Geggel, Robert L.


Publisher
John Wiley and Sons
Year
1997
Tongue
English
Weight
103 KB
Volume
40
Category
Article
ISSN
0098-6569

No coin nor oath required. For personal study only.

✦ Synopsis


Balloon atrial septostomy remains an important interventional procedure in the pediatric age group. The incidence of potential complications using the conventional balloon ranges from 0-11%. The purpose of this study was to evaluate a new low-profile end-hole septostomy balloon catheter with dual lumen, inserted via a 5F or 6F sheath. Seventeen neonates and infants with various forms of congenital heart disease requiring palliation underwent septostomy using the new catheter at a median age of 19 days (with a range of 1-593 days), and a median weight of 3.4 kg (a range of 2.5-8.4 kg).

The aortic saturation increased from 72 6 20% to 87 6 7%, p F .001; and the gradient across the atrial septum decreased (a-wave gradient from 11.2 6 10.3 to 2.1 6 3.6 mm Hg; v-wave from 10.4 6 7.7 to 1.2 6 1.2 mm Hg; and mean gradient from 8.5 6 6.9 to 0.9 6 1.3 mm Hg, p F .002). The diameter of the defect increased from 2.7 6 1.7mm to 8 6 2.3 mm, p F .001. There were no complications. We conclude that this new lowprofile septostomy catheter is safe and effective in creating a large defect size between the atria. Because of the smaller inflation size of the balloon and smaller introducer sheath compared with the conventional catheter, this new septostomy catheter should be especially useful in small neonates. Cathet.


πŸ“œ SIMILAR VOLUMES


Initial experience with the Europassβ„’: A
✍ Zimarino, Marco ;Corcos, Thierry ;Favereau, Xavier ;Tamburino, Corrado ;Toussain πŸ“‚ Article πŸ“… 1994 πŸ› John Wiley and Sons 🌐 English βš– 307 KB πŸ‘ 1 views

One of the causes for percutaneous transluminal coronary angioplasty (PTCA) failure is the inability to cross the lesion with the balloon catheter after guidewire positioning. The Europass@ coronary angioplasty catheter is a monorail Duralynm balloon catheter developed to enhance lesion crossability

A new thrombosuction catheter for corona
✍ van den Bos, Arjan A. ;van Ommen, Vincent ;Corbeij, Huub M. A. πŸ“‚ Article πŸ“… 1997 πŸ› John Wiley and Sons 🌐 English βš– 176 KB πŸ‘ 1 views

The presence of thrombus in a coronary vessel during percutaneous revascularisation can prevent adequate restoration of flow; it is also associated with an increased risk of distal embolization. We report the acute results and longer-term outcome of seven patients who underwent treatment with a new