✦ Synopsis


Objectives: To optimize the interpretation of GC-MS toxicological screenings (i.e., to facilitate ion specific queries, create custom reports specifically adapted to each confirmation procedure, and eliminate redundant and/or inaccurate data on library search reports).

Design and Methods:

The MS Chemstation software of the Hewlett Packard 5972 is constructed in a modular way. We made extensive modifications to two modules, the data analysis and the report modules, using the built-in MS Chemstation macro language.

Results: Ion specific queries were automated for over 60 commonly encountered analytes. Custom reports were created for the confirmation of positive drugs-of-abuse immunoassay results. With the incorporation of decision support rules into the data processing and the reporting phases, we obtained sensitive, accurate, and concise reports.

Conclusions: