✦ Synopsis


Abstract

Microcontroller‐based circuitry was built and tested for automatically tuning flexible RF receiver coils at the touch of a button. This circuitry is robust to 10% changes in probe center frequency, is in line with the scanner, and requires less than 1 s to tune a simple probe. Images were acquired using this circuitry with a varactor‐tunable 1‐inch flexible probe in a phantom and in an in vitro porcine knee model. The phantom experiments support the use of automatic tuning by demonstrating 30% signal‐to‐noise ratio (SNR) losses for 5% changes in coil center frequency, in agreement with theoretical calculations. Comparisons between patellofemoral cartilage images obtained using a 3‐inch surface coil and the surgically‐implanted 1‐inch flexible coil reveal a worst‐case local SNR advantage of a factor of 4 for the smaller coil. This work confirms that surgically implanted coils can greatly improve resolution in small‐field‐of‐view (FOV) applications, and demonstrates the importance and feasibility of automatically tuning such probes. Magn Reson Med, 2005. © 2005 Wiley‐Liss, Inc.


📜 SIMILAR VOLUMES


Automatic Tuned MRI RF Coil for Multinuc
✍ L.Tugan Muftuler; Gultekin Gulsen; Kumsal D. Sezen; Orhan Nalcioglu 📂 Article 📅 2002 🏛 Elsevier Science 🌐 English ⚖ 151 KB

We have developed an MRI RF coil whose tuning can be adjusted automatically between 120 and 128 MHz for sequential spectroscopic imaging of hydrogen and fluorine nuclei at field strength 3 T. Variable capacitance (varactor) diodes were placed on each rung of an eight-leg low-pass birdcage coil to ch

Examples of the design of screened and s
✍ Michael Burl; Ian R. Young 📂 Article 📅 1996 🏛 John Wiley and Sons 🌐 English ⚖ 460 KB

## Abstract Three methods of screening RF surface coil receivers are discussed. The aim is to design coils with very much greater sensitivity on one side than on the other so that signal and noise contributions from unwanted regions of the body can be minimized. These methods can usefully be applie

A multielement RF coil for MRI guidance
✍ Qiang Zhang; Michael Wendt; Andrik J. Aschoff; Jonathan S. Lewin; Jeffrey L. Due 📂 Article 📅 2001 🏛 John Wiley and Sons 🌐 English ⚖ 159 KB

## Abstract Accurate localization of minimally invasive devices is critical to the success of interventional procedures. Device orientation and tip position are two of the most important pieces of information needed to define device location for magnetic resonance imaging (MRI)‐guided interventiona

Integrated and interactive position trac
✍ Glyn A. Coutts; David J. Gilderdale; Ming Chui; Larry Kasuboski; Nandita M. Deso 📂 Article 📅 1998 🏛 John Wiley and Sons 🌐 English ⚖ 740 KB

## Abstract A method is described of tracking the position of a rigid device within an MR scanner and imaging with the image slice position determined by the current position of the device, such that the manipulation of the device can be achieved in a robust, interactive manner. The device can be e

Automatic tuning and matching of a small
✍ Rigoberto Pérez de Alejo; Carlos Garrido; Palmira Villa; Ignacio Rodriguez; Juan 📂 Article 📅 2004 🏛 John Wiley and Sons 🌐 English ⚖ 221 KB

## Abstract A new circuit design for automatically tuning and matching a saddle coil for small animal imaging is presented. This design allows working at ^1^H, ^19^F, and ^3^He resonance frequencies in a 4.7 T spectrometer. It is based on a balanced circuit with commercial variable capacity diodes