𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Automatic filtration and filter flush for robust online solid-phase extraction liquid chromatography

✍ Scribed by Kristin O. Svendsen; Hanne R. Larsen; Siri A. Pedersen; Ida Brenna; Elsa Lundanes; Steven R. Wilson


Book ID
102447639
Publisher
John Wiley and Sons
Year
2011
Tongue
English
Weight
101 KB
Volume
34
Category
Article
ISSN
1615-9306

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Online solid‐phase extraction‐liquid chromatography (SPE‐LC) with microbore or capillary columns was significantly improved regarding robustness, as an easily installed automatic filtration and filter flushing (AFFL) procedure was added to avoid system clogging. Specifically, an injected sample is passed through a union containing a stainless steel filter prior to SPE trapping. The filter stops any particulate matter from reaching the SPE. When the SPE is subsequently connected to the LC column by column switching, a separate pump backflushes the filter‐union, removing the particulate matter off the filter after each injection. This feature greatly reduced backpressure buildup over the entire system.