Automatic analysis of normal bone marrow blood cells using the XE-2100 automated hematology analyzer

โœ Scribed by Hisako Shibata; Takahisa Yamane; Ryousuke Yamamura; Kensuke Ohta; Takayuki Takubo; Tomio Kamitani; Masayuki Hino


Publisher
John Wiley and Sons
Year
2003
Tongue
English
Weight
133 KB
Volume
17
Category
Article
ISSN
0887-8013

No coin nor oath required. For personal study only.

โœฆ Synopsis


Abstract

The bone marrow aspiration test conventionally has been performed by visual methods, using a light microscope, because automatic blood cell analyzers cannot adequately capture erythroblasts and immature granulocytes (IGs) (Tatsumi et al.: Osaka City Med J 1988;34:135โ€“146; Tatsumi et al.: Am J Clin Pathol 1986;86:50โ€“54). With the development of the XEโ€2100 automatic blood cell analyzer (Sysmex Corporation, Kobe, Japan) in 1999, the classification of erythroblasts and IGs by means of flow cytometry (Zini et al.: Infus Ther Transfus Med 2001;28:277โ€“279; Briggs et al.: Sysmex J Int 1999;9:113โ€“119) became possible. In the present study we classified cells in 65 bone marrow aspiration specimens by the microscopic method and with the XE2100, and compared the results. A good correlation was found in the nucleated red blood cell (NRBC), white blood cell (WBC), and total nucleated cell (TNC) counts; the myeloid/erythroid ratio (M/E ratio); neutrophils, lymphocytes, eosinophils, and IGs in the immature myeloid information (IMI) channel; and the total cell count. These items can all be analyzed in about 54 sec with the XE2100, which is faster than with the microscopic method. Therefore, analysis of bone marrow aspiration specimens with this apparatus appears to be very useful for clinical screening as well as laboratory testing. J. Clin. Lab. Anal. 17:12โ€“17, 2003. ยฉ 2003 Wileyโ€Liss, Inc.


๐Ÿ“œ SIMILAR VOLUMES


Possible automatic cell classification o
โœ Ryousuke Yamamura; Takahisa Yamane; Masayuki Hino; Kensuke Ohta; Hisako Shibata; ๐Ÿ“‚ Article ๐Ÿ“… 2002 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 36 KB

## Abstract In clinical hematology, the demand for bone marrow aspiration testing is increasing. However, conventional automatic blood cell analyzers cannot completely analyze erythroblasts, and evaluation has mainly been performed by visual examination (the microscopic method). Using the CELLโ€DYN