Automated pre-column derivatization of thiolic fruit-antibrowning agents by sequential injection coupled to high-performance liquid chromatography using a monolithic stationary phase and an in-loop stopped-flow approach

✍ Scribed by Theano D. Karakosta; Paraskevas D. Tzanavaras; Demetrius G. Themelis


Publisher
John Wiley and Sons
Year
2011
Tongue
English
Weight
176 KB
Volume
34
Category
Article
ISSN
1615-9306

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

The present study reports the very first application of ethyl propiolate (EP) as an advantageous pre‐column derivatization reagent for the determination of thiols by liquid chromatography (LC). Cysteine (CYS), glutathione (GSH) and N‐acetylcysteine (NAC) were derivatized online under stopped‐flow conditions in a sequential injection (SI) system coupled to HPLC. The formed derivatives were separated isocratically with a monolithic stationary phase (100×4.6 mm id) and UV detected at 285 nm. Critical parameters that affected the online pre‐column derivatization reaction (e.g. the reaction time and the amount concentration of EP) and the separation (e.g. pH and the composition of the mobile phase) were investigated. The developed analytical scheme offers a total analysis time of less than 10 min, limits of detection in the range of 0.24–0.35 μmol/L and satisfactory linearity up to 200 μmol/L for all analytes. The proposed method was applied to the analysis of the selected thiols – that are often employed as antibrowning agents – in fresh fruit samples.