Automated determination of molybdenum(VI) in seawater by means of constant-current reduction of the adsorbed 8-quinolinol complex in a computerized flow potentiometric stripping analyzer

✍ Scribed by Chi Hua; Daniel Jagner; Lars Renman


Publisher
Elsevier Science
Year
1987
Tongue
English
Weight
370 KB
Volume
192
Category
Article
ISSN
0003-2670

No coin nor oath required. For personal study only.

✦ Synopsis


Molybdenum(V1) in seawater is determined by means of potentiostatic adsorption of the 8quinolinol complex onto a mercury film electrode at -0.2 V vs. SCE and subsequent reduction of the complex by means of constant-current stripping in 5 M calcium chloride medium with a fully automated stripping analyzer. A single stripping peak at -0.42 V vs. SCE was obtained. The molybdenum(V1) concentration in reference seawater NASS-1, with a certified value of 11.5 + 1.9 pg 1-l , was found to be 8.9 f 1.3 /.lg 1-l (n = 10).