Automated detection of local artery wall thickness based on M-line signal processing

✍ Scribed by Arnold P.G. Hoeks; Christine Willekes; Pierre Boutouyrie; Peter J. Brands; Jean M. Willigers; Robert S. Reneman


Publisher
Elsevier Science
Year
1997
Tongue
English
Weight
839 KB
Volume
23
Category
Article
ISSN
0301-5629

No coin nor oath required. For personal study only.

✦ Synopsis


The Young's modulus of an arterial segment, a measure of the elastic properties of the arterial wall, requires the simultaneous and local assessment of pulse pressure, wall thickness, diameter, and distensibility (relative increase in cross-sectional area per change in blood pressure). The diameter and relative increase in cross-sectional area can be obtained with a wall track system, processing the radiofrequency (r.f.) ultrasound signals received along a single line of observation (M-line processing). It will be demonstrated that it is feasible to combine, in a single measurement, the assessment of wall thickness and the (relative change in) diameter involving a minimum of user interaction. Phantom tests show a standard error of the estimate for intima-media thickness measurements of less than 20 microns; in vivo registrations exhibit a variation on the order of 45 microns. It is concluded that processing of the radiofrequency ultrasound signal, acquired along an M-line, provides an accurate and time-efficient alternative for videoprocessing of 2-dimensional B-mode ultrasound images to estimate artery wall thickness.