𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Approaching the 4He lambda line with a liquid nitrogen precooled two-stage pulse tube refrigerator

✍ Scribed by G. Thummes; S. Bender; C. Heiden


Publisher
Elsevier Science
Year
1996
Tongue
English
Weight
368 KB
Volume
36
Category
Article
ISSN
0011-2275

No coin nor oath required. For personal study only.

✦ Synopsis


We report on the design and test of a three-stage refrigerator, consisting of a liquid nitrogen bath with heat exchangers and two subsequent pulse tube stages. The liquid nitrogen bath serves for precooling the oscillating 4He-gas flow. The PTR stages are operated in orifice and double-inlet mode with the orifice valves, reservoirs and second-inlet valves located at ambient temperature. Er,Ni shot is used for the regenerator matrix in the coldest stage. A minimum temperature of 2.13(l) K has been achieved, which, at the average operating pressure of 19.5 bar, is only 0.19 K above the h-line for the superfluid phase transition of 4He.