๐”– Bobbio Scriptorium
โœฆ   LIBER   โœฆ

Application of the generalized internal reference method for the characterization of parameters causing drift in inductively-coupled plasma emission spectrometry

โœ Scribed by Avraham Lorber; Zvi Goldbart


Publisher
Elsevier Science
Year
1984
Tongue
English
Weight
756 KB
Volume
161
Category
Article
ISSN
0003-2670

No coin nor oath required. For personal study only.

โœฆ Synopsis


Drift in signals limits the performance of inductively-coupled plasma (1.c.p.) emission spectrometry. The generalized internal reference method was applied to deconvolute the drift of an i,c.p. system. The plasma parameters were correlated to the drift components and the error boundary for the method was found. In the system studied, it was found that fluctuations in the nebulizer uptake rate were responsible for short-term signal changes. Long-term signal drift was caused mainly by flow variations of both cooling gas and aerosol carrier gas. The incident power to the plasma was found to be stable compared to other sources of drift. The method was found to be adequate for evaluating the origins of i.c.p. drift.