Apparent transducer non-reciprocity in an ultrasonic flow meter

✍ Scribed by Jan van Deventer; Jerker Delsing


Book ID
104263516
Publisher
Elsevier Science
Year
2002
Tongue
English
Weight
155 KB
Volume
40
Category
Article
ISSN
0041-624X

No coin nor oath required. For personal study only.

✦ Synopsis


This paper investigates the effects of non-identical ultrasonic transducers on reciprocity and zero-flow calibration in transit time flow meters. According to the theorem of reciprocity, there should not be any difference between the up- and downstream acoustic times of flight in a zero-flow situation. This would thus eliminate zero-flow estimation drifts. The flow meter is modeled as a one dimensional system with equivalent electrical circuits and simulated with simulation program with integrated circuits emphasis. The work shows that variations between the two transducers cause false estimates of flow and indicate which parameters have the largest influence. It indicates that reciprocity holds only for identical transducers.