Anion separations with pressure-assisted capillary electrophoresis using a sequential injection analysis manifold and contactless conductivity detection

✍ Scribed by Thanh Duc Mai; Peter C. Hauser


Publisher
John Wiley and Sons
Year
2011
Tongue
English
Weight
293 KB
Volume
32
Category
Article
ISSN
0173-0835

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

It is demonstrated that a hydrodynamic flow superimposed on the mobility of analyte anions can be used for the optimization of analysis time in capillary zone electrophoresis. It was also possible to use the approach for counter‐balancing the electroosmotic flow and this works as well as the use of surface modifiers. To avoid any band‐broadening due to the bulk flow narrow capillaries of 10 μm internal diameter were employed. This was enabled by the use of capacitively coupled contactless conductivity detection, which does not suffer from the downscaling, and detection down to between 1 and 20 μM for a range of inorganic and small organic anions was found feasible. Precisely controlled hydrodynamic flow was generated with a sequential injection manifold based on a syringe pump. Sample injection was carried out with a new design relying on a simple piece of capillary tubing to achieve the appropriate back‐pressure for the required split‐injection procedure.