𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Analytical error propagation in diffusion anisotropy calculations

✍ Scribed by Aziz Hatim Poonawalla; Xiaohong Joe Zhou


Publisher
John Wiley and Sons
Year
2004
Tongue
English
Weight
351 KB
Volume
19
Category
Article
ISSN
1053-1807

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Purpose

To develop an analytical formalism describing how noise and selection of diffusion‐weighting scheme propagate through the diffusion tensor imaging (DTI) computational chain into variances of the diffusion tensor elements, and errors in the relative anisotropy (RA) and fractional anisotropy (FA) indices.

Materials and Methods

Singular‐value decomposition (SVD) was used to determine the tensor variances, with diffusion‐weighting scheme and measurement noise incorporated into the design matrix. Anisotropy errors were then derived using propagation of error. To illustrate the applications of the model, 12 data sets were acquired from each human subject, over a range of b‐values (500–2500 seconds/mm^2^) and diffusion‐weighting gradient directions (N = 6–55). The mean RA and FA values and their respective errors were calculated within a region of interest (ROI) in the splenium. The RA and FA errors as a function of b‐value and N were evaluated, and a number of diffusion‐weighting schemes were assessed based on a new metric, sum of diffusion tensor variances.

Results

When the acquisition time was held constant, the sum of the diffusion tensor variances decreased as N increased. The same trend was also observed for several diffusion‐weighting schemes with constant condition number when noise in the diffusion‐weighted (DW) images was assumed unity. Errors in both FA and RA increased with b‐value and decreased with N. The FA error in the splenium was approximately threefold smaller than RA error, irrespective of b‐value or N.

Conclusion

The condition number may not adequately characterize the noise sensitivity for a given diffusion‐weighting scheme. Signal averaging may not be as effective as increasing N, especially when N is small (e.g., N < 13). Due to its smaller error, FA is preferred over RA for quantitative DTI applications. J. Magn. Reson. Imaging 2004;19:489–498. © 2004 Wiley‐Liss, Inc.


📜 SIMILAR VOLUMES


Does fractional anisotropy have better n
✍ Khader M. Hasan; Andrew L. Alexander; Ponnada A. Narayana 📂 Article 📅 2004 🏛 John Wiley and Sons 🌐 English ⚖ 207 KB

## Abstract Fractional anisotropy (FA) and relative anisotropy (RA) are the two most commonly used scalar measures of anisotropy in diffusion tensor (DT) MRI. While a few published studies have shown that FA has superior noise immunity relative to RA, no theoretical basis has been proposed to expla

Calculation of zinc diffusion profiles i
✍ Eiji Kawazura; Yoshiharu Shimose; Hiroaki Yoshino; Kenji Kawano 📂 Article 📅 2006 🏛 John Wiley and Sons 🌐 English ⚖ 494 KB

In this article we use the implicit finite-difference beam propagation method (implicit FD-BPM) to calculate acceptor density profiles created by the diffusion of Zn into InP, InGaAs, and InGaAsP. We verify the accuracy of this method by comparing our computed results with carrier density profiles m