✦ Synopsis


A multitechnique approach has been used to characterize the surface of nonosseointegrated titanium implants and the surrounding biological tissues. Five pure titanium dental implants were used as reference, and 25 removed implants were studied. Surface and in-depth chemical compositions of the implants (from a total of 16 patients) were investigated by X-ray photoelectron spectroscopy (XPS). Histological slides of the surrounding tissues were examined by light microscopy, XPS, and electron microprobe analysis. None of the failed implants presented the regular surface composition and depth profile of the TiO 2 overlayer; foreign elements (Ca, Na, P, Si, Cl, Zn, Pb, and Al) were observed on some implants. Fibrosis, lymphocytic and plasmocytic infiltrates, and granulomatous lesions were detected in the surrounding tissues. XPS and electron microprobe analysis indicated the presence of Zn, Fe, Sn, and Ti in the tissues. As a possible scenario for implant failure, we propose and discuss a oxidoreduction mechanism, leading to a partial dissolution or the complete dissociation of the protective titanium dioxide overlayer and to ion diffusion through the surrounding tissues.