𝔖 Bobbio Scriptorium
✦   LIBER   ✦

An optosensor for riboflavin with C18 silica gel as a substrate

✍ Scribed by Zhilong Gong; Zhujun Zhang


Publisher
Elsevier Science
Year
1997
Tongue
English
Weight
377 KB
Volume
339
Category
Article
ISSN
0003-2670

No coin nor oath required. For personal study only.

✦ Synopsis


A flow-through optosensor for the sensitive determination of riboflavin in some pharmaceutical preparations was developed. The sensor was developed in conjunction with a flow-injection analysis system and uses C 18 silica gel as a substrate. The wavelengths of excitation and emission were 466 and 515 nm, respectively. The applicable concentration range was 2.2-91.8 ng ml-' with a detection limit (304terion) of 0.4 ng ml-'. The relative standard deviation for ten determinations of 1.5 ng ml-' of riboflavin was 1.1%. The method is more sensitive than other methods described in the literature.