𝔖 Bobbio Scriptorium
✦   LIBER   ✦

An inductively-coupled, detachable receiver coil system for use with magnetic resonance compatible endoscopes

✍ Scribed by David J. Gilderdale; Andreanna D. Williams; Umakant Dave; Nandita M. deSouza


Publisher
John Wiley and Sons
Year
2003
Tongue
English
Weight
450 KB
Volume
18
Category
Article
ISSN
1053-1807

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Purpose

To construct an inductively‐coupled receiver coil system for use with a magnetic resonance (MR) compatible endoscope, and to evaluate its use in a pilot group of patients with esophageal cancer.

Materials and Methods

An inductively‐coupled coil system, comprising a saddle geometry cylindrical receiver coil fitted as a sleeve around the endoscope tip and a pick‐up coil housed within a channel of an MR‐compatible endoscope, was designed and developed for use at 0.5 T. Twenty‐three patients with esophageal cancer were recruited for MR endoscopy. In 17 cases, the endoscopic coil system was used in conjunction with an external surface coil in order to obtain information from the surrounding mediastinum. The examination took 40–50 minutes.

Results

MR imaging using the inductively‐coupled endoscopic coil was successful in 21 cases (one failed intubation and one artifact from unrelated external source). Image artifact was largely due to respiration and global patient motion in sedated individuals undergoing endoscopy.

Conclusion

Inductively‐coupled coil systems may be used with endoscopes to allow improved safety through increased patient‐system isolation and detachability of coils and electronics for repair or replacement with coils tuned for different frequencies. J. Magn. Reson. Imaging 2003;18:131–135. © 2003 Wiley‐Liss, Inc.