An apparatus for accurate determinations of vapour-liquid equilibrium properties and gas PVT properties
โ Scribed by Aa. Fredenslund; J. Mollerup; L.J. Christiansen
- Publisher
- Elsevier Science
- Year
- 1973
- Tongue
- English
- Weight
- 399 KB
- Volume
- 13
- Category
- Article
- ISSN
- 0011-2275
No coin nor oath required. For personal study only.
โฆ Synopsis
A precision apparatus for the experimental determination of vapour-liquid equilibrium properties and PVT properties of gas mixtures has been constructed and thoroughly tested. The apparatus used in the phase equilibrium determinations is of the vapour recirculation type, and that used in the PVT measurements is a Burnett apparatus.
The experiments may be performed in the temperature range-180 to +25ยฐC, the temperature being controlled to the nearest +-0.01ยฐC. The temperature can be measured to the nearest 0.01 K using a high precision ac bridge and a calibrated resistance thermometer. The pressure range is from 3 to 350 atm (1 atm = 101.3 kN m "2) on the vapour recirculation apparatus and up to 700 arm on the Burnett apparatus; the pressure is determined with a dead weight piston gauge, which has an accuracy of 0.01% of the reading or 0.005 Ib in "2, whichever is the greatest. The composition of the gas and liquid'phases is determined using a carefully calibrated gas chromotograph. Both the gas and the liquid samples are injected directly into the carrier gas of the gas chromatograph.