Amperometric detection of amino acids in a flow-injection system with a nickel(II)-modified electrode with an Eastman-AQ polymer film

✍ Scribed by Anhua Liu; Erkang Wang


Publisher
Elsevier Science
Year
1993
Tongue
English
Weight
588 KB
Volume
280
Category
Article
ISSN
0003-2670

No coin nor oath required. For personal study only.

✦ Synopsis


Amperometic flow measurements were made at +0.55 V (vs. Ag/AgClI in 0.1 mol 1-t KOH electrolyte with an Ni(I1) chemically modified electrode (CME) with an Eastman-AQ polymer film. The use and characteristics of a Ni(II)-containing crystalline and polymer-modified electrode obtained by a double coating step as a detector for amino acids in a flow-injection system using reversed-phase liquid chromatography are described. The detection of these analytes is based on the higher oxidation state of nickel (NiOOH) controlled by the applied potential. The electroanalytical parameters and the detection current for a series of amines and amino acids were investigated. The use of such a CME in the flow-injection technique was found to be suitable in a solution at low PH. The linear range for glycine is 5 X lo-'-0.1 mol 1-l with a detection limit of 1.0 X 10e6 mol I-'. A 1 X 10m4 mol 1-r mixture of serine and tyrosine was also detected after separation on an Nucleosil Crs column.