Flow reversal and flow recycle configurations in an air-segmented flow injection system are presented for simultaneous determinations based on differential kinetics with a slow enzymatic reaction. The sensitivity of determination by the proposed system is improved over that by flow-reversal flow inj
Air-segmented flow injection: a hybrid technique for automated, low dispersion determinations
โ Scribed by Yunsheng Hsieh; S.R. Crouch
- Publisher
- Elsevier Science
- Year
- 1995
- Tongue
- English
- Weight
- 589 KB
- Volume
- 303
- Category
- Article
- ISSN
- 0003-2670
No coin nor oath required. For personal study only.
โฆ Synopsis
A novel approach to the design and use of an air-segmented flow injection (ASFI) system is described. The ASFI system possesses features of both classical flow injection (FI) and air-segmented continuous flow (ASCF) methods. The entire air-segmented sample zone is injected through a simple 6-way valve into the carrier stream. Preliminary studies of dye dispersion and spectrophotometric glucose determinations in wine and serum samples are presented. These results indicate that the proposed system has advantages for most determinations in sensitivity, sampling rate and solution consumption when compared to the other flow analysis techniques. Analytical results are also compared with those obtained by kinetic-based determinations.