## Abstract ## Purpose To evaluate the performance of a real‐time MR system for interventional procedures that adjusts specific image parameters in real time based on a catheter's speed of insertion. ## Materials and Methods The system was implemented using only the hardware provided with a stan
Active catheter tracking using parallel MRI and real-time image reconstruction
✍ Scribed by Michael Bock; Sven Müller; Sven Zuehlsdorff; Peter Speier; Christian Fink; Peter Hallscheidt; Reiner Umathum; Wolfhard Semmler
- Publisher
- John Wiley and Sons
- Year
- 2006
- Tongue
- English
- Weight
- 546 KB
- Volume
- 55
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
In this work active MR catheter tracking with automatic slice alignment was combined with an autocalibrated parallel imaging technique. Using an optimized generalized autocalibrating partially parallel acquisitions (GRAPPA) algorithm with an acceleration factor of 2, we were able to reduce the acquisition time per image by 34%. To accelerate real‐time GRAPPA image reconstruction, the coil sensitivities were updated only after slice reorientation. For a 2D trueFISP acquisition (160 × 256 matrix, 80% phase matrix, half Fourier acquisition, TR = 3.7 ms, GRAPPA factor = 2) real‐time image reconstruction was achieved with up to six imaging coils. In a single animal experiment the method was used to steer a catheter from the vena cava through the beating heart into the pulmonary vasculature at an image update rate of about five images per second. Under all slice orientations, parallel image reconstruction was accomplished with only minor image artifacts, and the increased temporal resolution provided a sharp delineation of intracardial structures, such as the papillary muscle. Magn Reson Med, 2006. © 2006 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES
A real-time NMR image reconstruction and display system was developed using a high-speed personal computer and optimized for the 32-bit multitasking Microsoft Windows 95 operating system. The system was operated at various CPU clock frequencies by changing the motherboard clock frequency and the pro
## Abstract ## Purpose: To measure contrast agent enhancement kinetics in the liver and to further evaluate and develop an optimized gadolinium enhanced MRI using a single injection real‐time bolus‐tracking method for reproducible imaging of the transient arterial‐phase. ## Materials and Methods:
## Abstract A novel two‐element, catheter‐based phased array coil was designed and built for both active MR device tracking and high‐resolution vessel wall imaging. The device consists of two independent solenoid coils that are wound in opposite directions, connected to separate receive channels, a
## Abstract A previously proposed nonlinear inverse reconstruction for autocalibrated parallel imaging simultaneously estimates coil sensitivities and image content. This work exploits this property for real‐time MRI, where coil sensitivities need to be dynamically adapted to the conditions generat