𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Absolute 60Co characterization based on gamma–gamma coincident detection by two NaI(Tl) detectors

✍ Scribed by Peter Volkovitsky; Philip Naudus


Book ID
103857846
Publisher
Elsevier Science
Year
2009
Tongue
English
Weight
298 KB
Volume
607
Category
Article
ISSN
0168-9002

No coin nor oath required. For personal study only.

✦ Synopsis


A gamma-gamma coincidence measurement method was used for the characterization of 60 Co. A twodimensional analysis of coincident events allows the careful calculation of the number of counts, when each of the two g-rays emitted by 60 Co is in the photopeak area in each of the NaI(Tl) detectors. The standard formulas that were applied for absolute calibration of 125 I (Eldridge-Crowther formulas) were modified for the case of 60 Co decay. These modified Eldridge-Crowther formulas were applied to the absolute calibration of 60 Co samples measured by the two NaI(Tl) detectors both in coincidence and normal modes. Measurements were performed for three sources of different activities and for two positions of each source between the two NaI(Tl) detectors. Results are in good agreement with the measurements of the same samples done by the 4pb-g method.


📜 SIMILAR VOLUMES