Abstract

Magnetic field monitoring with NMR probes has recently been introduced as a means of measuring the actual spatiotemporal magnetic field evolution during individual MR scans. Receive‐only NMR probes as used thus far for this purpose impose significant practical limitations due to radiofrequency (RF) interference with the actual MR experiment. In this work these limitations are overcome with a transmit/receive (T/R) monitoring system based on RF‐shielded NMR probes. The proposed system is largely autonomous and protected against RF contamination. As a consequence the field probes can be positioned freely and permit monitoring imaging procedures of arbitrary geometry and angulation. The T/R approach is also exploited to simplify probe manufacturing and remove constraints on material choices. Probe miniaturization permits monitoring imaging scans with nominal resolutions on the order of 400 μm. The added capabilities of the new probes and system are demonstrated by first in vivo results, obtained with monitored gradient‐echo and spin‐echo echo‐planar imaging (EPI) scans. Magn Reson Med, 2009. © 2009 Wiley‐Liss, Inc.


📜 SIMILAR VOLUMES


2H transmit–receive NMR probes for magne
✍ Pekka Sipilä; Sebastian Greding; Gerhard Wachutka; Florian Wiesinger 📂 Article 📅 2011 🏛 John Wiley and Sons 🌐 English ⚖ 356 KB

Measuring image encoding fields in real time and applying the information in postprocessing offer improved image quality for MRI, particularly for applications that are intrinsically sensitive to gradient imperfections. For this task, a stand-alone magnetometer system based on multiple 2 H transmit-

Evaluation of MRI for in vivo monitoring
✍ Laura Pop-Fanea; Sonia Nielles Vallespin; James M. S. Hutchison; John V. Forrest 📂 Article 📅 2004 🏛 John Wiley and Sons 🌐 English ⚖ 557 KB

## Abstract Two quantitative methods were developed for investigation of the potential of MRI for in vivo monitoring of retinal damage and detachment in experimental autoimmune uveitis (EAU). Measurements of retinal thickness and detachment area were performed on matched MR and histologic (HIST) im

In vivo cellular imaging of magnetically
✍ Pierre Smirnov; Florence Gazeau; Maïté Lewin; Jean Claude Bacri; Nathalie Siauve 📂 Article 📅 2004 🏛 John Wiley and Sons 🌐 English ⚖ 289 KB

The feasibility of in vivo cellular imaging using a 1.5 T clinical magnet was studied in the mouse. Hybridoma cells were labeled with anionic gamma-Fe2O3 superparamagnetic iron oxide nanoparticles. These were internalized by the endocytose pathway. Both electron spin resonance and magnetophoresis as

MR monitoring of laser-induced lesions o
✍ Delphine Germain; Patrick Chevallier; Alexandre Laurent; Michael Savart; Michel 📂 Article 📅 2001 🏛 John Wiley and Sons 🌐 English ⚖ 301 KB 👁 2 views

The aims of this study were, firstly, to monitor temperature with magnetic resonance (MR) during laser ablations performed in pig livers in vivo in a low-field open scanner (0.23T) and, secondly, to study the feasibility of lesion size prediction. Spin-echo (SE) images of 29 sec acquired during lase