Calibration of flow cytometers is becoming an increasingly important issue for both quality control of instrument performance and quantitation of antibody binding capacity of cells. Due to the numerous different instruments and analysis software currently available, a standardized method of calibrat
A standard for calibration and shading correction of a fluorescence microscope
β Scribed by Michael A. Model; Janis K. Burkhardt
- Publisher
- John Wiley and Sons
- Year
- 2001
- Tongue
- English
- Weight
- 142 KB
- Volume
- 44
- Category
- Article
- ISSN
- 0196-4763
No coin nor oath required. For personal study only.
β¦ Synopsis
Background: Numerous applications of fluorescence microscopy require quantitation of signal intensity in reproducible units. Two problems must be overcome to achieve this goal. First, due to various instrumental factors, the same sample imaged on two microscopes or even on the same microscope at different times may produce highly divergent readings. Second, because of shading, some areas within the same field may appear brighter than others despite the same amount of fluorophore. The first type of variability requires calibration using a sample of reproducible fluorescence yield; to correct for shading, a uniform fluorescent field is needed. Methods: Standard slides were prepared by placing several microliters of 10%-50% w/v fluorescein or rhodamine between a coverglass and a slide. They were used to perform shading correction and normalization under a variety of imaging conditions.
Results: Concentrated fluorophores produced a uniform fluorescent field of moderate and reproducible brightness. By expressing the staining of a biological object in the units of standard slides, identical results were obtained irrespective of the imaging conditions or the microscope used. We compared shading correction based on concentrated fluorescein with two other standards. Concentrated fluorescein resulted in the best equalization of the field. Conclusions: Standardization of fluorescent images can be achieved by normalizing them to the image of a concentrated solution of a fluorophore. Due to its simplicity and efficiency, this method can be used in clinical analysis as well as in routine laboratory practice. Cytometry 44: 309 -316, 2001.
π SIMILAR VOLUMES
In the development of suitable standards and calibration materials for fluorescence measurement, it becomes necessary to make accurate fluorescence measurements of these materials on flow cytometers. The results of such measurements may be affected by numerous sources of error, prominent among which
## Abstract Calibration is required for most soil moisture sensors if accurate measurements are to be obtained. This can be time consuming and costly, especially if field calibration is undertaken, but can be facilitated by a good understanding of the behaviour of the particular sensor being calibr
## Abstract Standardization, calibration, and controls (negative and positive controls) are essential for quality assurance. Cytometers are capable of reliable and repeatable cellular analyses. However, a prerequisite is instrument calibration and standardized preanalytics. Calibration is often don