A technique is described for preventing interference of sorbitol with the assay of P, by modifying the procedure of B. N. Ames (1966, in Methods in Enzymology, E. F. Neufeld and V. Ginsburg, eds., Vol. 8, pp. 115-118, Academic Press, New York). The new method relies on the ability of precipitated pr
A simple automated method for determination of low concentrations of inorganic phosphate
โ Scribed by E. Ehrlich; R. Coelho; F. Guillain
- Publisher
- Elsevier Science
- Year
- 1972
- Tongue
- English
- Weight
- 244 KB
- Volume
- 50
- Category
- Article
- ISSN
- 0003-2697
No coin nor oath required. For personal study only.
๐ SIMILAR VOLUMES
We present here an improvement in the classical molybdate method for inorganic phosphate determination. This method has high sensitivity, with 1 nmol Of Pi giving an absorbance change of 0.060 &a unit. It is highly reproducible and the color remains stable for at least 3.5 h. In addition the use of
A microprocedure for the colorimetric determination of inorganic pyrophosphate (PP0 in the presence or absence of orthophosphate (Pi) has been developed. PPi is estimated quantitatively as the amount of chromophore formed with molybdate reagent, 1-amino-2-naphthol-4-sulfonic acid in bisulfite and th
The inherently simple biuret reaction is hampered by lack of sensitivity and reproducibility in the determination of low protein concentrations. A method utilizing a modified biuret formulation and colorimeter range expander permits protein determination in the 0.01-3.0 mg/ml range at a sampling rat