𝔖 Bobbio Scriptorium
✦   LIBER   ✦

A sensitive high vacuum gauge using an electron multiplier

✍ Scribed by EE Windsor


Publisher
Elsevier Science
Year
1970
Tongue
English
Weight
246 KB
Volume
20
Category
Article
ISSN
0042-207X

No coin nor oath required. For personal study only.

✦ Synopsis


In the conventional Bayard-A/pert ion gauge undesirable interactions occur between the gauge and its environment due to fhe pumping action which is inherent in the operation of the gauge, and to the desorption and dissociation of gas by heated components.

These effecfs can be minimized by using an electron multiplier as an ion detector and amplifier, and the necessary small ionizing current can be provided by a cold emitter.

A practical form of gauge having a digital output and providing a linear response to pressure changes over three orders of magnitude is described.