A novel system of four-dimensional motion analysis after total hip arthroplasty

✍ Scribed by Keisuke Hagio; Nobuhiko Sugano; Takashi Nishii; Hidenobu Miki; Yoshito Otake; Asaki Hattori; Naoki Suzuki; Kazuo Yonenobu; Hideki Yoshikawa; Takahiro Ochi


Publisher
Elsevier Science
Year
2004
Tongue
English
Weight
655 KB
Volume
22
Category
Article
ISSN
0736-0266

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

We have developed a novel system of four‐dimensional motion analysis after total hip arthroplasty (THA) that can aid in preventing dislocation by assessing safe range of motion for patients in several daily activities.

This system uses skeletal structure data from CT and motion capture data from an infrared position sensor. A 3‐D model reconstructed from CT data is combined with the motion capture data. Using this system, we analyzed hip motion when getting up from and sitting down in a chair or picking up an object while sitting in a chair in 17 patients (26 hips) who underwent THA. To assess the accuracy of this system's measurements, open MRI was used to evaluate positions of skin markers against bones in five healthy volunteers in various postures.

No impingement between bones and/or implants was found in any subjects during any activities. However, mean angle at the point of maximum hip flexion was different for each patient. The open MRI results indicated that average error in hip angle of the present system was within 5Β° for each static posture.

The functional position of the pelvis during daily activities must be taken into account when assessing the real risk of dislocation. The present system enables dynamic analysis involving not only alignment of components and bones of each patient, but also individual differences in characteristics of daily motions. Further investigation using this system can help determine safe ranges of motion for preventing hip dislocation, improving the accuracy of individualized guidance for patients regarding postoperative activities. Β© 2003 Orthopaedic Research Society. Published by Elsevier Ltd. All rights reserved.


πŸ“œ SIMILAR VOLUMES


Bony impingement affects range of motion
✍ Oliver Kessler; Shantanu Patil; Wirth Stefan; Eckart Mayr; Clifford W. Colwell J πŸ“‚ Article πŸ“… 2008 πŸ› Elsevier Science 🌐 English βš– 32 KB

The name of the third listed author of this article has been spelled incorrectly as well as transposed. His name is Stephan Wirth. The publisher very much regrets this error. The article to which this Erratum refers appears in Volume 26, Number 4, pages 443-452.

Bony impingement affects range of motion
✍ Oliver Kessler; Shantanu Patil; Wirth Stefan; Eckart Mayr; Clifford W. Colwell J πŸ“‚ Article πŸ“… 2008 πŸ› Elsevier Science 🌐 English βš– 523 KB

## Abstract Hip range of motion after total hip arthroplasty has been shown to be dependent on prosthetic design and component placement. We hypothesized that bony anatomy would significantly affect range of motion. Computer models of a current generation hip arthroplasty design were virtually impl