A novel microbial BOD biosensor developed by the immobilization of P. Syringae in micro-cellular polymers
✍ Scribed by Serdar Kara; Bülent Keskinler; Elif Erhan
- Publisher
- Wiley (John Wiley & Sons)
- Year
- 2009
- Tongue
- English
- Weight
- 270 KB
- Volume
- 84
- Category
- Article
- ISSN
- 0268-2575
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
BACKGROUND: A biochemical oxygen demand (BOD) sensor, based on an immobilized Pseudomonas syringae in highly porous micro‐cellular polymer (MCP) in combination with a dissolved oxygen electrode, has been developed for the analysis of biodegradable organic compounds in aqueous samples. Microorganisms were immobilized in a molded MCP disk and a wastewater sample was injected into the biocomposite disk by a flow injection system. Dissolved oxygen (DO) changes as a measure of soluble BOD was read with a DO probe placed into a flow cell carrying biocatalytically activated disk.
RESULTS: Optimal response of the MCP BOD sensor was obtained at pH 6.8 and 25 °C with a typical response time of 3–5 min for a 2 mm thick molded polymeric disk. The sensor showed detection linearity over the range 5–100 mg L^−1^ BOD~5~ (r^2^ > 0.99) at a flow rate of 0.6 mL min^−1^. The repeatability and reproducibility of the sensor response were found to be 3.08% and 7.77%, respectively. BOD values produced with this biosensor for various municipal and industrial wastewaters correlated well with those determined by the conventional 5‐day BOD test.
CONCLUSION: This new biosensor was different from present amperometric BOD biosensor configurations in which the biocatalyst (microbial/enzymatic) is placed between cellulose and Teflon membranes installed on a DO probe. The use of a molded MCP disk coniainng microbial activity offers better stability and lifetime for commercial use in environmental monitoring. Copyright © 2008 Society of Chemical Industry