A modified quadrupole gradient set for use in high resolution MRI tagging
โ Scribed by Walter G. O'Dell; Joseph S. Schoeniger; Stephen J. Blackband; Elliot R. McVeigh
- Publisher
- John Wiley and Sons
- Year
- 1994
- Tongue
- English
- Weight
- 579 KB
- Volume
- 32
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
โฆ Synopsis
Abstract
A special purpose gradient set with a Cos(20) current distribution has been constructed for high resolution MR imaging of small samples in a clinical Signa 1.5 T scanner using the existing gradient amplifiers. The X, Y, and Z gradient coils can attain gradient field strengths of 11.3, 4.7, and 15.2 G/cm at 100 amps current, respectively, with a slew rate of 20 G/cm/ms and usable ramp time of 150 ฮผs. Field distortions are less than 2% over the central 8 cm of the bore, suitable for high resolution tagging of isolated canine hearts.