A Miniature Cylindrical Quadrupole Ion Trap: Simulation and Experiment
โ Scribed by Badman, Ethan R.; Johnson, Rudolph C.; Plass, Wolfgang R.; Cooks, R. Graham
- Book ID
- 126999908
- Publisher
- American Chemical Society
- Year
- 1998
- Tongue
- English
- Weight
- 79 KB
- Volume
- 70
- Category
- Article
- ISSN
- 0003-2700
No coin nor oath required. For personal study only.
๐ SIMILAR VOLUMES
Two small (5 mm internal radius) cylindrical ion traps (CITs) are arranged in series and operated using a single ion source, detector and radio frequency (rf) trapping signal. Ions are trapped in the first CIT and later transferred to the second by applying a direct current (dc) pulse to the endcap
This paper is a presentation of recent key developments in quadrupole ion trap mass spectrometry. These key developments have been made in three areas; namely, the trapping of ions generated externally to the ion trap, the fine control of the trajectories of ions confined within the ion trap, and th