𝔖 Bobbio Scriptorium
✦   LIBER   ✦

A method for preamplifier-decoupling improvement in quadrature phased-array coils

✍ Scribed by Ceylan Celil Guclu; Ed Boskamp; Tsinghua Zheng; Ricardo Becerra; LeRoy Blawat


Publisher
John Wiley and Sons
Year
2004
Tongue
English
Weight
219 KB
Volume
19
Category
Article
ISSN
1053-1807

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Purpose

To improve preamplifier decoupling in a quadrature phased‐array coil using a new quadrature combiner configuration.

Materials and Methods

A commercial six‐element quadrature cervical‐thoracic‐lumbar (CTL) coil was used on a 1.5 T GE Signa Lx Scanner. The new combiner configuration is designed based on the fact that when one compares the amount of flux linkage of two loops with butterfly elements separated by the same distance, the loop elements have much higher coupling than the butterfly elements.

Results

The results are demonstrated using an imaging‐based element coupling measurement method. The new quadrature combiner configuration has introduced an approximately −8dB improvement in preamplifier decoupling.

Conclusion

One‐dimensional quadrature phased arrays can be built even if the preamplifier is not integrated in the coil. If the combiner is placed between the coil and the preamplifiers, the butterflies should be on the high impedance side of the quadrature combiner. The loops should see the low input impedance of the preamplifiers. J. Magn. Reson. Imaging 2004;19:255–258. © 2004 Wiley‐Liss, Inc.


📜 SIMILAR VOLUMES


Quadrature coil design for high-resoluti
✍ Mohamed Ouhlous; Adriaan Moelker; Herman J. Flick; Piotr A. Wielopolski; Thomas 📂 Article 📅 2007 🏛 John Wiley and Sons 🌐 English ⚖ 503 KB

## Abstract ## Purpose To evaluate the ability of a custom‐built coil design to provide improved signal‐to‐noise ratio (SNR) and less signal drop with increasing depth at the carotid artery. ## Materials and Methods Phased‐array surface coils can provide a high SNR to image the carotid vessel wa

Improvements in carotid plaque imaging u
✍ Niranjan Balu; Vasily L. Yarnykh; Joshua Scholnick; Baocheng Chu; Chun Yuan; Cec 📂 Article 📅 2009 🏛 John Wiley and Sons 🌐 English ⚖ 392 KB

## Abstract ## Purpose To design and compare an eight‐channel phased array (PA) coil for carotid imaging to an established four‐channel PA design at 3T. ## Materials and Methods An eight‐channel PA (8PA) coil was designed specifically for imaging the carotid bifurcation and compared with the exi