New perspectives for epithermal neutron spectroscopy have been opened up as a result of the development of the Resonance Detector and its use on inverse geometry time-of-flight spectrometers at spallation sources. A special application of the Resonance Detector is the Very Low Angle Detector Bank (V
A high-resolution two-dimensional 3He neutron MSGC with pads for neutron scattering experiments
โ Scribed by K. Fujita; H. Takahashi; P. Siritiprussamee; H. Niko; K. Nishi; Y. Takada; T. Oku; J. Suzuki; T. Ino; H.M. Shimizu; M. Furusaka
- Publisher
- Elsevier Science
- Year
- 2007
- Tongue
- English
- Weight
- 594 KB
- Volume
- 580
- Category
- Article
- ISSN
- 0168-9002
No coin nor oath required. For personal study only.
โฆ Synopsis
High-resolution neutron imaging detectors are required for future high-intensity neutron sources such as the J-PARC project in Japan. We are developing a high-resolution gaseous neutron detector by using the microstrip gas chamber (MSGC) with floating-pad readout method. In this paper, we describe the neutron detector performance obtained with this MSGC and the readout method. For a 3 cm ร 3 cm MSGC filled with 5 bar CF 4 and 0.5 bar 3 He, we could achieve a measured position resolution of 0.6 mm (FWHM) including the beam size and the parallax error.