๐”– Bobbio Scriptorium
โœฆ   LIBER   โœฆ

A fully automated contour detection algorithm the preliminary step for scatter and attenuation compensation in SPECT

โœ Scribed by R. Ben Younes; J. Mas; R. Bidet


Book ID
104682708
Publisher
Springer
Year
1988
Tongue
English
Weight
372 KB
Volume
14
Category
Article
ISSN
0340-6997

No coin nor oath required. For personal study only.

โœฆ Synopsis


Contour detection is an important step in information extraction from nuclear medicine images. In order to perform accurate quantitative studies in single photon emission computed tomography (SPECT) a new procedure is described which can rapidly derive the best fit contour of an attenuated medium. Some authors evaluate the influence of the detected contour on the reconstructed images with various attenuation correction techniques. Most of the methods are strongly affected by inaccurately detected contours. This approach uses the Compton window to redetermine the convex contour: it seems to be simpler and more practical in clinical SPECT studies. The main advantages of this procedure are the high speed of computation, the accuracy of the contour found and the programme's automation. Results obtained using computer simulated and real phantoms or clinical studies demonstrate the reliability of the present algorithm.