A comparison of anaesthetic tensions in arterial blood and oxygenator exhaust gas during cardiopulmonary bypass

✍ Scribed by G. G. Lockwood; S. M. Sapsed-Byrne; S. Adams


Publisher
John Wiley and Sons
Year
1999
Tongue
English
Weight
129 KB
Volume
54
Category
Article
ISSN
0003-2409

No coin nor oath required. For personal study only.

✦ Synopsis


This study evaluates the usefulness of the analysis of gas sampled from the exhaust port of a membrane oxygenator in the estimation of anaesthetic tension in arterial blood. Sixty‐seven arterial blood samples were drawn from patients undergoing hypothermic cardiopulmonary bypass with anaesthesia maintained by either isoflurane or desflurane. Anaesthetic tensions in the oxygenator exhaust gas were measured using an infrared analyser and in arterial blood using a two‐stage headspace technique with a gas chromatograph. Both measurement systems were calibrated with the same standard gas mixtures. There was no difference in anaesthetic tension measured in arterial blood and gas leaving the oxygenator exhaust (isoflurane: n = 29, range: 0.3–0.8%, 95% limits of agreement: − 0.08% to 0.09%; desflurane: n = 38, range: 1.5–5.4%; 95% limits of agreement − 0.65% to 0.58%). We conclude that anaesthetic tensions in arterial blood can be accurately monitored by analysis of the gas emerging from the exhaust port of a membrane oxygenator.