## Abstract The effect of using ammonia as a carrier gas on the response of the flame ionization detector (FID) has been investigated. It was found that the FID response, calculated as the effective carbon number (__ECN__), increased for all the compounds studied when ammonia, rather than helium, w
Fast response flame ionisation detectors react to small combustible particles in the gas stream as well as to ionisable gases, producing unwanted large, short duration spikes in the output signal. During some wind tunnel dispersion modelling experiments, this problem arose due to sampling the atmosp
A typical, pellet-type alkali flame detector was monitored photometrically. Its predominant (but not exclusive) mode of response is negative, i.e. alkali emission decreases when gas chromatographic effluents are burned in the flame. As a detector for organic halides, the optical emission of the alka